SERVICES PROVIDED FOR COMPOUND, SPECIALTY
AND METALLURGICAL MICROSCOPES
Per
ACC Proprietary Procedure # MP-FFCAL and # NPWI-OPTCLN
·
Perform
full evaluation of equipment.
·
Full
cleaning of exterior surfaces.
· Clean all optical elements including eyepieces, beam split mirror, objectives and final magnification lens.
·
Check
stage operation.
·
Lubricate
all applicable areas using synthetic lube.
· Check operation of objective turret. Make minor adjustments if necessary.
·
Check all
mechanical hardware for looseness - tighten as needed.
·
Clean and
adjust illumination for maximum output and best image clarity.
·
Clean,
center, focus and adjust photo and/or video attachments.
·
Perform
minor mechanical adjustments to optimize system.
·
Calibrate
all objectives to installed reticle or video measurement system as applicable.
·
Furnish
Certificate of Accuracy with Before/After data.
·
If
necessary, furnish written report detailing additional problems found.
Note any action needed to increase system productivity.
Note estimated costs regarding needed actions (if available).
·
ACC glass
master stage micrometer calibrated traceable to NIST to .00001”
(10
microinches)
·
ACC steel
and ceramic master gage blocks calibrated traceable to NIST to .000002”(2 microinches)
·
All
calibrations conform to ANSI/NCSL Z540-1 and meet ISO requirements.
ABSOLUTE CLARITY & CALIBRATION,
LLC
109 Main
Street v
Terryville, CT
06786
(860)583-0502 v
(860)314-1851 (fax)